The Hardware Test screen is displayed when the self-verification test is complete. The Hardware
Test screen displays the functions that pass and fail.
4.
(Optional) Tap Export to export the results.
5.
Tap Close.
Use Smart Help if any hardware functions fail (see “Request technical support with Smart Help” on
page 172).
View the last hardware test result
The self verification test checks the instrument hardware functions. You can view the results of the last
self-verification test that was run.
1.
In the home screen, tap
(Settings)4Maint
enance and service4Self verification test.
The Self-Verification Test screen is displayed.
2.
Tap Hardware test.
3.
Tap Last test.
The Self-Verification Test screen displays the functions that pass and fail.
Use Smart Help if any hardware functions fail (see “Request technical support with Smart Help” on
page 172).
Perform a peripheral test
The peripheral test takes approximately 5 minutes. Tap
(Close) on any scr
een to exit the peripheral
test.
The peripheral test checks the following functions:
Touchscreen accuracy
Speaker
Microphone
Camera
Proximity sensor
1.
In the home screen, tap
(Settings)4Maint
enance and service4Self verification test.
The Self Verification Test screen is displayed.
2.
Tap Peripheral test.
3.
Tap Start.
The peripheral test begins with the touch accuracy test.
Chapter 11 Configur
e the instrument settings
Configure maintenance and service settings
11
174
QuantStudio
6 Pro Real-Time PCR System and QuantStudio
7 Pro Real-Time PCR System User Guide