Anritsu MT1000A - Test Results

Anritsu MT1000A
605 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Evaluation itemEvaluation item
Use the drop-down menu to select the item to evaluate. If selecting other than
Any Alarm or ErrorAny Alarm or Error, another menu appears.
Evaluation typeEvaluation type
Select the relevant type.
Pass & fail thresholdPass & fail threshold
Touch the left-hand number to set the lower limit for "Warning". Touch the
right-hand number to set the lower limit for "Fail". The lower limit of "Fail"
must be equal to or greater than the lower limit of "Warning" (defining a
"Within limits" range in between them).
77..33..33 Test ResultsTest Results
77..33..33..11 SummarySummary
Touching the SummarySummary button in the navigation area will display the screen
shown below.
This screen shows the summary of all error count and ratio. You can insert the
pattern error in this screen.
BERBER
Displays the result of pattern errors and threshold.
Statistics CategoryStatistics Category
The lamp icon in Status column shows the Pass or Fail results. Touching the
Status column displays the statistics results.
TransportTransport
Displays the result of Pass & fail threshold if Transport is selecting in Test SetupTest Setup
screen.
PatternPattern
BERTBERT
437437

Table of Contents

Other manuals for Anritsu MT1000A

Related product manuals