EasyManua.ls Logo

Anritsu MT8820A - Initialization Procedures

Anritsu MT8820A
296 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5.1 Initialization
5-3
5.1 Initialization
Before starting a new measurement, perform initialization. Initialization returns all
the parameters to the known state to prevent from being affected by the previous op-
erations.
The procedure for initialization is as follows.
[Procedure]
1. Press
Preset
to open the Preset menu.
2. Press
F1
: Preset to execute the initialization.
Example of programming by remote control
*RST
Initialization by execution of Preset menu or via GPIB interface initializes the fol-
lowing items:
• Selection screen
• All parameters on the measurement screen
• Call processing status
• Measurement status/results display
The following items are not initialized:
• Setting for the System Configuration screen
• Calibration data acquired by executing Band Calibration
• Calibration data acquired by executing Full Calibration

Table of Contents