5.1 Initialization
5-3
5.1 Initialization
Before starting a new measurement, perform initialization. Initialization returns all
the parameters to the known state to prevent from being affected by the previous op-
erations.
The procedure for initialization is as follows.
[Procedure]
1. Press
Preset
to open the Preset menu.
2. Press
F1
: Preset to execute the initialization.
Example of programming by remote control
*RST
Initialization by execution of Preset menu or via GPIB interface initializes the fol-
lowing items:
• Selection screen
• All parameters on the measurement screen
• Call processing status
• Measurement status/results display
The following items are not initialized:
• Setting for the System Configuration screen
• Calibration data acquired by executing Band Calibration
• Calibration data acquired by executing Full Calibration