Section 7 Remote Control Using PC
7-28
7.6.9 ERR event status enable register
The ERR event status enable register is a register consisting of eight bits.
Each bit of the ERR event status enable register corresponds to that of the ERR status
register. Logical OR of the bits is output to bit 3 of the status byte register.
That is, when any status specified by the ERR event status enable register has oc-
curred, bit 3 of the status byte register becomes 1 (true).
Setting
Set the END event status enable register by using the ESE3 command. The
parameter is the decimal number converted from the sum of the values obtained
by binary-weighing the specified bits.
Example:
To set bit 3 of the status byte register to 1 when both the sending measurement
error and receiving measurement error have occurred:
ESE3 12 /*Errors correspond to bit 2 and bit 3, so 2
2
+ 2
3
= 12 is set.*/
Query
Read using the ESE3? query. Response is the decimal number converted from
the sum of the values obtained by binary-weighing the specified bits.
Clear
The following methods are available to clear the ERR event status enable register:
Send the ESE3 0 command