Section 8 Remote Control Commands
8-66
SWP
Measure - Single (Synchronization)
Starting a measurement in the synchronous single mode
Function
Starts a measurement in the single mode. Processing for the next command does
not start before completion of the measurement, allowing synchronization between
MT8820A and controller.
Query inquires whether or not the measurement has been completed.
Command
SWP
Query
SWP?
Response
status
Parameters
status Measurement status
0 : Measurement completed
1 : Measurement being performed
Sample programming
To start measurement in synchronous single mode:
SWP
Related commands
TS : Same function as SWP (no query)
SNGLS : Starting measurement in the single mode
S2 : Starting measurement in the single mode