Section 5 Using Common Functions
5-6
5.3 Starting a Measurement
A measurement is performed on the measurement screen (RF-1 Screen). A meas-
urement cannot be performed on the Common Screen.
Common screen on which a measurement cannot be performed
• System Configuration screen
• System Information screen
• Parameter Save screen
• Parameter Recall screen
5.3.1 Measurement mode
The following three measurement modes are available:
(1) Single mode
A measurement is performed only once. Averaging measurements are made
by the specified number of averaging times (number of measurements).
(2) Continuous mode
A measurement in Single mode is repeated unless the measurement stop condi-
tion is met by pressing
Stop
, etc.
(3) Synchronous single mode (for remote control only)
This mode is only for remote control. A measurement is performed in the sin-
gle mode. If the measurement is started with SWP command, the next com-
mand is not processed until the measurement finishes. This can prevent from
returning the response of the measurement result before the measurement com-
pletion.
5.3.2 Starting a measurement in the Single mode
[Procedure]
Press
Single
on the measurement screen.
Example of programming by remote control
SNGLS
5.3.3 Starting a measurement in the Continuous mode
[Procedure]
Press
Continous
on the measurement screen.
Example of programming by remote control
CONTS