Section 8 Remote Control Commands
8-18
*SRE
Service Request Enable Command
Setting service request enable register
Function
Sets the service request enable register. When the event set in the service request
enable register occurs, the MSS bit (bit 6) of the status byte register becomes “1”
(true).
Command
*SRE register
Query
*SRE?
Response
register
Parameters
register Service request enable register
Value = bit0 + bit1 + ... + bit7
bit0 = 2
0
= 1 : Not used
bit1 = 2
1
= 2 : Not used
bit2 = 2
2
= 4 : ERR event status register
bit3 = 2
3
= 8 : END event status register
bit4 = 2
4
= 16 : MAV
bit5 = 2
5
= 32 : ESB
bit6 = 2
6
= 64 : MSS or RQS
bit7 = 2
7
= 128 : Not used
Description
From among the values 2
0
= 1, 2
1
= 2, 2
2
= 4, 2
3
= 8, 2
4
= 16, 2
5
= 32, 2
6
= 64, and 2
7
= 128, which correspond to bits 1, 2, 3, 4, 5, 6 and 7 of the service request enable
register; the sum of values of bits desired to be enabled becomes the parameter.
Sample programming
To set MSS bit to “1” (true) when an end event occurs:
*SRE 8