Section 8 Remote Control Commands
8-30
ESE2
END Event Status Enable Register
Setting end event status enable register
Function
Sets the end event status enable register. When the set end-event occurs, the end
summary bit (ESB, bit 2) of the status byte register becomes “1” (true).
Command
ESE2 register
Query
ESE2?
Response
register
Parameters
register End event status enable register
Value = bit0 + bit1 + ... + bit7
When using W-CDMA or GSM measurement software
bit0 = 2
0
= 1 : Measurement/sweep complete
bit1 = 2
1
= 2 : Not used
bit2 = 2
2
= 4 : Reception measurement complete
bit3 = 2
3
= 8 : Transmission measurement complete
bit4 = 2
4
= 16 : Audio measurement complete
bit5 = 2
5
= 32 : Not used
bit6 = 2
6
= 64 : Not used
bit7 = 2
7
= 128 : Not used
When using CDMA2000 measurement software
bit0 = 2
0
= 1 : All measurements complete
bit1 = 2
1
= 2 : Access Probe Power measurement complete
bit2 = 2
2
= 4 : Reception measurement complete
bit3 = 2
3
= 8 : Transmission measurement complete (excepting Access
Probe Power)
bit4 = 2
4
= 16 : Not used
bit5 = 2
5
= 32 : Not used
bit6 = 2
6
= 64 : Not used
bit7 = 2
7
= 128 : Not used