EasyManua.ls Logo

Anritsu MT8820A - Page 185

Anritsu MT8820A
296 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7.6 Status Report
7-27
Table 7-6-8-2 Description of each bit of ERR event status register
When using CDMA2000 measurement software
Bit Description
7 Not used
6 Call Drop Error
Occurs when the MT8820A can’t demodulate signal of phone.
5 Not used
4 Call Processing Error
Indicates that an error has occurred during call processing.
3 Reception Measurement Error
Indicates that an error has occurred during reception measurement.
2
Transmission Measurement Error
Indicates that an error has occurred during transmission measurement,
except for Access Prove Power.
1 Access Prove Power Measurement Error
Indicates that an error has occurred during measurement for Access
Prove Power on the Fundamental Measurement screen.
0 Not used
Query
Read using the ESR3? query. When readout is successful, the contents of the
register are cleared.
Example:
Query ESR3?
Response 80
(80 = 2
4
+ 2
6
: bit 4 and bit 6 errors have occurred)
Clear
The following methods are available to clear the ERR status register:
Read the status using the command
Send the *CLS command
Turn the power On