This procedure applies to a:
Standard 384-Well Block
Standard 96-Well Block
Fast 96-Well Block
1. Remove the sample block from the 7900HT instrument as explained in
“Removing the Sample Block” on page 7-7.
2. Use the following figure as a guide to locate the suspected contaminated wells
on the sample block.
3. Pipette the appropriate volume of 10% bleach solution into each suspected
contaminated well of the sample block module.
For a Standard 384-Well Block, pipette 40 µL bleach solution to each well.
For a Standard 96-Well Block, pipette 150 µL bleach solution to each well.
For a Fast 96-Well Block, pipette 55 µL bleach solution to each well.
4. Allow the sample block to sit for 3 to 5 min.
5. Using a pipet, remove the bleach solution from the wells of the sample block.
6. Rinse (pipette and remove) each contaminated well with three treatments of
deionized water at the appropriate volume for the sample block.
For a Standard 384-Well Block, rinse affected wells with 40 µL deionized
water.
For a Standard 96-Well Block, rinse affected wells with 150 µL deionized
water.
For a Fast 96-Well Block, rinse affected wells with 55 µL deionized water.
Note: Absolute isopropanol can be substituted for water in the third treatment.
7. Remove any remaining isopropanol or water from the wells of the sample block
module.
8. Replace the sample block as explained in “Replacing the Sample Block” on
page 7-10.
9. Run a background plate to confirm that the contamination has been removed.
Warning
GR2028
Well A-1
Circuitry and connections to
the instrument (Do Not Touch)

Table of Contents

Other manuals for Applied Biosystems 7900HT

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7900HT and is the answer not in the manual?

Applied Biosystems 7900HT Specifications

General IconGeneral
BrandApplied Biosystems
Model7900HT
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals