EasyManua.ls Logo

Applied Biosystems 7900HT - Page 263

Applied Biosystems 7900HT
384 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DRAFT
September 1, 2004 11:39 am, CH_Maintenance.fm
Verifying Instrument Performance
Applied Biosystems 7900HT Fast Real-Time PCR System and SDS Enterprise Database User Guide 7-31
Purpose of
TGF-β Runs
The TaqMan
®
Low Density 7900HT Installation Array (TGF-β card) is an
experiment run to verify the performance of the 7900HT instrument for use with
TaqMan Low Density Arrays. The TGF-β card is preloaded with the reagents
necessary for the detection and quantification of genomic copies of the TGF-β gene.
Each well contains preloaded TGF-β primers and FAM MGB dye-labeled probe.
Materials
Required
RNase P Runs
Appropriate TaqMan RNase P Instrument Verification Plate (see Table 7-2 on
page 7-30)
Centrifuge, with plate adaptor
TGF-β Runs
•TGF-β card
•TaqMan
®
2 Universal PCR Master Mix
Water, DNase/RNase-free, deionized
2-mL microtube, sterile
Centrifuge
Microfluidic card sealer
Taq Ma n
®
RNase P
Instrument Verification
Plate
(PN 4310982, standard
96-well plate)
Table 7-2 Configurations of the TaqMan RNase P Instrument Verification Plates
TaqMan RNase P
Instrument Verification
Plate
Sample Configuration
GR2108
Unknown 1
5000
NTC STD 1250 STD 2500
STD 5000 STD 10000 STD 20000
Unknown 2
10000

Table of Contents

Other manuals for Applied Biosystems 7900HT

Related product manuals