September 1, 2004 11:39 am, CH_Trouble.fm
TaqMan Low Density Array
Applied Biosystems 7900HT Fast Real-Time PCR System and SDS Enterprise Database User Guide 8-19
After pipetting, too little
PCR reaction mixture has
gone into the fill reservoir.
The PCR reaction mixture
was not correctly pipetted
into the fill reservoir.
Care must be taken to
correctly pipette the PCR
reaction mixture (100 µL)
into the fill reservoir.
1. Carefully follow the
procedures provided in
“Loading the TaqMan
Low Density Arrays” on
page 4-14.
2. After centrifuging, be
sure to note the fluid
levels in the fill reservoir
as described in step 6 on
page 4-18.
After pipetting, some of the
PCR reaction mixture leaks
out of the vent port in the
fill reservoir.
The PCR reaction mixture
leaks during PCR cycling.
The aluminum foil backing
may have been damaged
during the sealing step.
Discard the Low Density
Array.
Note: See “About the
Sealer” on page 4-14 for
proper sealing instructions.
After a mean assay result
and standard deviation of
the replicates are
calculated, a specific
replicate appears to be an
outlier.
N/A Review the multicomponent
analysis display for that
replicate. Delete the outlier
and reanalyze.
Table 8-6 Troubleshooting the TaqMan
®
Low Density Array
Observation Possible Cause Recommended Action

Table of Contents

Other manuals for Applied Biosystems 7900HT

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Biosystems 7900HT and is the answer not in the manual?

Applied Biosystems 7900HT Specifications

General IconGeneral
BrandApplied Biosystems
Model7900HT
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals