Alarm Display Chapter 9 - Alarms
Apollo Operator’s Manual 135
Part Number: 90 38 237, 6th edition
Alarm Display
Figure 101. Multiple Alarms DisplayAlarm messages are displayed in the alarm field in
order of priority. Up to three messages can be
displayed simultaneously. In some cases, the
corresponding measured values are highlighted on
the screen by a flashing background in addition to
the alarm message.
If more than three alarms occur simultaneously, the
symbol “ “ appears to the right of the alarm field
(1 in Figure 101), and the soft key >Show All
Alarms< is activated on the right side of the screen
(2 in Figure 101). When this soft key is pressed, the
upper curve display is replaced by up to six
additional alarm fields for 15 seconds (see Figure
102).
The curve reappears when the soft key >Show All
Alarms< is pressed again or when the 15 seconds
have expired.
Figure 102. Additional Alarm Fields