Chapter 11 - Cleaning and Maintenance Apollo Cleaning and Disinfection
172 Apollo Operator’s Manual
Part Number: 90 38 237, 6th edition
Removing the passive scavenger system (optional):
Figure 143. Passive Scavenger System
1. Remove the waste gas hose connected between
the hospital waste gas disposal system and the
bottom connection on the scavenger (1 in Figure
143).
2. Remove the scavenger hose connected between
the scavenger and the scavenger connection on
the back of the anesthesia machine (2 in Figure
143).
3. Remove the scavenger system from the
machine.
4. Prepare the parts for cleaning and disinfection.
Note: For cleaning/disinfection instructions for the
passive scavenger system, refer to its
Instructions for Use.
Figure 144. Suction System Bracket and RegulatorRemoving the suction system (optional)
1. Remove the suction bottle from the slide mount
(1 in Figure 144).
2. Remove the suction regulator from the bracket (2
in Figure 144).
3. Remove the suction system bracket from the
side rail on the side of the anesthesia machine.
4. If using disposable suction bottle, dispose of the
bottle with infectious waste.
5. If using reusable suction bottle, empty the bottle
and dispose of contents with infectious waste.
Note: For cleaning/disinfection instructions for the
reusable suction bottle, refer to its
Instructions for Use.
Note: For cleaning/disinfection instructions for the
suction regulator, refer to its Instructions for
Use.
1
2
2
1
2
1
WARNING !
Always wear gloves when emptying suction
bottle.