Configuration During Operation Chapter 10 - Configuration
Apollo Operator’s Manual 161
Part Number: 90 38 237, 6th edition
Volumes/Alarms
Figure 128. Volumes/Alarms Configuration ScreenThe following settings can be selected in the
Volumes/Alarms configuration screen (refer to Figure
128).
Alarm Volume
- 1 is minimum volume*
- 9 is maximum volume
*Settings 1 through 3 are not available per U.S.
standards..
Note: The alarms NO O
2
DELIVERY and NO O
2
SUPPLY are always annunciated at the
maximum volume, regardless of the setting.
Breathing Sound
- 0 is off
- 9 is maximum volume
The breathing sound is produced by the Breathing
Sound Emulator (BSE) module. This module
transforms the measured inspiratory and expiratory
flow values into an acoustic tone similar to a
breathing sound. The sound volume is dependent on
the set patient age in order to produce optimal
volume at all flow levels.
All alarms are still audible with the breathing sound
volume set to the maximum.
Pulse Volume (for SpO
2
, optional)
- 0 is off
- 9 is maximum volume
Alarms On/Off
This setting is used to enable/disable CO
2
alarms,
optional SpO
2
alarms, and bypass mode (see
page 139 and page 140 for complete information).
WARNING !
Ensure that the alarm volume is set to a
sufficiently loud level.