EasyManuals Logo

Dräger Apollo Operating Instructions

Dräger Apollo
232 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #65 background imageLoading...
Page #65 background image
Overview Chapter 5 - Preuse Checkout
Apollo Operator’s Manual 57
Part Number: 90 38 237, 6th edition
Overview
The preuse checkout procedure must be performed
to ensure that the Apollo is ready for use. This is a
recommended procedure. Follow the institution’s
policies for specific procedures.
If the Apollo fails any checkout routine, do not use
the machine until corrective action is taken. If
indicated, contact an authorized representative of
DrägerService for inspection of the unit.
Checking the Workstation
According to the Check List
The preuse checkout procedure consists of a manual
procedure performed by the user, followed by an
automated self test. The manual procedure is
summarized in the check list that is displayed after
the machine is powered on.
Before starting the checkout procedure, ensure the
following conditions are met:
the machine is prepared and assembled as
specified in the Routine Maintenance and
System Setup sections of this manual
the gas supply and power supply are
connected
WARNING !
Do not insert any additional components into or
modify the Apollo after the checkout procedure
has been started.
WARNING !
Patient hoses should be adjusted to the
appropriate lengths prior to performing the leak
and compliance tests. Inappropriate hose
length affects compliance and can result in
incorrect tidal volume delivery to the patient.

Table of Contents

Other manuals for Dräger Apollo

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Apollo and is the answer not in the manual?

Dräger Apollo Specifications

General IconGeneral
BrandDräger
ModelApollo
CategoryMedical Equipment
LanguageEnglish

Related product manuals