Chapter 13 - Specifications Specifications
212 Apollo Operator’s Manual
Part Number: 90 38 237, 6th edition
Breathing System
Auxiliary Common Gas Outlet (CGO)
(Optional)
Total gas volume (without breathing hoses, incl.
absorber)
in Man. Spont typically 4.4 L
in automatic mode typically 4.7 L
(incl. piston
volume)
Compliance (without breathing hoses)
in Man. Spont typically 5.3 mL/
cmH
2
O
in automatic mode typically 2.4 mL/
cmH
2
O
Absorber volume 1.5 L
Total system leakage (as per EN 740) <150 mL/min at
30 cmH
2
O
Pressure limitation valve APL
Adjustment range 5 to 70 cmH
2
O
Accuracy between 5 and 15 L/min ±15% of measured value or ±3 cmH
2
O, whichever is
greater
Resistance (as per ISO 8835-2, dry conditions) Absorber in normal operation
(filled with Drägersorb
®
800 Plus)
Drägersorb
®
CLIC, in normal
operation
(filled with Drägersorb
®
800 Plus)
Inspiratory Expiratory Inspiratory Expiratory
at 5 L/min -4.6 cmH
2
O 4.0 cmH
2
O -4.5 cmH
2
O 4.2 cmH
2
O
at 10 L/min -4.3 cmH
2
O 4.2 cmH
2
O -4.0 cmH
2
O 4.3 cmH
2
O
Minimal Limited Pressure (as per ISO 8835-5) -3 cmH
2
O
Connection Dia 22mm ISO cone (male) with dia. 15mm ISO cone (female)
Pressure limitation max. 80 cmH
2
O at 18 L/min
Fresh-gas flow 0 to 18 L/min
(tolerances as in “Fresh Gas Data”, see
page 211)

Table of Contents

Other manuals for Dräger Apollo

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Apollo and is the answer not in the manual?

Dräger Apollo Specifications

General IconGeneral
BrandDräger
ModelApollo
CategoryMedical Equipment
LanguageEnglish

Related product manuals