Chapter 13 - Specifications Specifications
210 Apollo Operator’s Manual
Part Number: 90 38 237, 6th edition
Operating Data
Electrical Data
Operating voltage 100 to 127 VAC, 50-60 Hz, 12.4 A max.
Power input 200 W typically, max. 1.5 kW with power drawn from auxiliary outlets
Uninterruptible power supply Fully charged batteries: at least 30 minutes;
Up to 90 minutes, depending on ventilation parameters
(auxiliary outlets are not powered)
Auxiliary outlets 2 outlets with automatic circuit breakers rated at 4 A each,
1 outlet for desflurane vaporizer rated 2 A; combined current for all outlets
is 6A.
The outlets are isolated against mains to reduce leakage current.
Connection for optional halogen lamp 12 V max. 10 W
Compressed gas supply at pipeline inlet
O
2
39 psi to 100 psi at max. 78 L/min (including max. output flow)
N
2
O 39 psi to 100 psi at max. 18 L/min
Air 39 psi to 100 psi at max. 18 L/min
Dew point >41°F below ambient temperature
Oil content <0.1 mg/m
3
Particles Dust-free air (filtered with pores <1 µm)
Driving gas consumption None
Noise emission in normal operation with
ventilation (breathing sound set to OFF)
<50 dB(A)
Protection class
Workstation I, in accordance with IEC 601-1
SpO
2
sensor Type BF
electrically isolated from protective conductor
Electromagnetic compatibility (EMC) Tested to IEC 601-1-2
m

Table of Contents

Other manuals for Dräger Apollo

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Apollo and is the answer not in the manual?

Dräger Apollo Specifications

General IconGeneral
BrandDräger
ModelApollo
CategoryMedical Equipment
LanguageEnglish

Related product manuals