Specifications Chapter 13 - Specifications
Apollo Operator’s Manual 219
Part Number: 90 38 237, 6th edition
Interfaces
CLIC Adapter Technical Data
2 serial interfaces: COM1, COM2
Protocol MEDIBUS
1
(COM 2 without real-time data)
Plug connector 9-pin sub-D, electrical isolation, 1.5 kV
Pin assignment
1 NC, not connected
2 TX, transmit
3 RX, receive
4 DTR, data terminal ready
5 GND, ground
6 DSR, data set ready
7 RTS, request to send
8 CTS, clear to send
9 NC, not connected
Shields DTR and DSR, as well as RTS and CTS are internally connected.
Hardware handshake is not supported.
Settings 1200 or 9600 Baud
even parity
8 data bits
1 stop bit
Dräger Base IV system (not sold in the U.S.) For connecting IV systems
SpO
2
For connecting an SpO
2
sensor
1
typical delay of system alarms: 600 ms
Resistance Reduced by approx. 1 mbar in the bypass mode, depending on the
breathing system of the anesthetic workstation
Compliance
Normal mode:
Bypass mode:
typically 1.3 mL/mbar
typically 0.4 mL/mbar
Leakage: <5 mL/min