Specifications Chapter 13 - Specifications
Apollo Operator’s Manual 211
Part Number: 90 38 237, 6th edition
Fresh Gas Data
Auxiliary O
2
Flowmeter
Settings:
O
2
concentration 21 to 100 Vol.%
Fresh-gas flow, dosing 0 to at least 12 L/min per gas (O
2
, N
2
O, Air)
Fresh-gas flow, electronic measuring 0 to 12 L/min volumetric flow per gas (O
2
, N
2
O, Air)
Accuracy ±10% or 0.12 L/min, whichever is greater
Resolution 0.01 L/min (from 0 to 0.2 L/min)
0.02 L/min (from >0.2 to 0.5 L/min)
0.05 L/min (from >0.5 to 1.0 L/min)
0.1 L/min (from >1 to 12 L/min)
Fresh-gas flow, total flowmeter 0 to 10 L/min
Accuracy ±10% full scale for 50% O
2
/ 50% N
2
O
calibrated at 5 L/min
Resolution 0.5 L/min (from 0.5 to 2 L/min)
1.0 L/min (from 2 to 10 L/min)
O
2
flush >35 L/min
Connection Staged connector for use with different hose diameters
Fresh-gas flow 0 to 10 L/min
Accuracy ± 5% of full scale
Resolution 0.5 L/min