EasyManua.ls Logo

Dräger Apollo

Dräger Apollo
232 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 12 - Troubleshooting Fresh Gas Delivery Failure
194 Apollo Operator’s Manual
Part Number: 90 38 237, 5th edition
Fresh Gas Delivery Failure
Figure 162. Gas Mixer Failure
If the fresh gas mixer fails, the following message is
displayed in the status field on the screen: “GAS
MIXER FAIL” and the numerical values for the fresh-
gas flows appear grayed out to indicate that they
may be inaccurate.
The current ventilation mode remains active.
Close N2O and air flow control valves
Check vaporizer setting
Set O2 flow to the desired level using the total
flow meter
Note: Refer to the total flow meter for approximate
flow (see the Specifications section of the
manual for accuracy).

Table of Contents

Other manuals for Dräger Apollo

Related product manuals