Chapter 12 - Troubleshooting Display Failure
196 Apollo Operator’s Manual
Part Number: 90 38 237, 5th edition
Display Failure
If there is no display on the monitor screen:
Switch the machine off.
•Set O
2
flow to the desired level using the total
flow meter.
Check the vaporizer setting.
Ventilate patient by hand.
Provide adequate substitute monitoring.
User Interface Failure
If the keypads, confirm knob, or flow control knobs
are not operational:
Select the monitoring mode (see page 126).
Ventilate patient by hand.
Observe the total flow meter for approximate
flow (see the Specifications section of the
manual for accuracy).
System Failure
If the system no longer responds to an action:
1. Ventilate the patient by hand.
2. Switch the machine off and on again.
3. Cancel the self test.
4. Check the vaporizer setting.
5. Close all flow controls (except O
2
), and run
100% O
2
.
WARNING !
If the breathing bag does not fill with fresh gas
or manual ventilation is not possible:
check the oxygen supply; open cylinder
valves if necessary
If fresh gas is still not delivered:
close all flow controls
disconnect the patient from the
workstation and use another device
turn the workstation off

Table of Contents

Other manuals for Dräger Apollo

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Apollo and is the answer not in the manual?

Dräger Apollo Specifications

General IconGeneral
BrandDräger
ModelApollo
CategoryMedical Equipment
LanguageEnglish

Related product manuals