Chapter 13 - Specifications Specifications
220 Apollo Operator’s Manual
Part Number: 90 38 237, 6th edition
AGS Scavenger
TechnicalData of AGS Scavenger
Ambient conditions
During operation:
Temperature 59 °F to 86 °F
Atmospheric pressure 525 to 795 mmHg
Rel. humidity 0 to 98%, without condensation
During storage:
Temperature –4 °F to 158 °F
Atmospheric pressure 375 to 825 mmHg
Rel. humidity 0 to 98%, without condensation
Performancedata
Effect on the anesthetic system/ventilator:
Increase in exhalation <0.5 cmH
2
O at 30 L/min
resistance <3.6 cmH
2
O at 75 L/min
Generated vacuum pressure <1 Pa (0.01 cmH
2
O)
Generated flow <50 mL/min
Effect on the environment:
Gas loss,
dependent on fresh gas flow & ventilation parameters <25 mL/min typical)
Operating data
Operating capacity of scavenging flow 30 to 50 L/min
Setting with service flow indicator 65 to 75 L/min
Particle filter pore width 0.1 mm
Particle filter resistance 5.1 cmH
2
O at 50 L/min
Supply performance requirements
across a flow resistance of 20 cmH
2
O min. 25 L/min suction
f
across a flow resistance of 10 cmH
2
O max. 50 L/min suction
Dimensions (WxHxD) 4.7in.x13.8 in. x 7.9 in.
Weight, without hoses approx. 2.1 lbs.

Table of Contents

Other manuals for Dräger Apollo

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Apollo and is the answer not in the manual?

Dräger Apollo Specifications

General IconGeneral
BrandDräger
ModelApollo
CategoryMedical Equipment
LanguageEnglish

Related product manuals