Specifications Chapter 13 - Specifications
Apollo Operator’s Manual 221
Part Number: 90 38 237, 6th edition
Latex Use
The Apollo is latex-free. Latex-free breathing bags
and breathing hoses must be used for latex-free use.
Electromagnetic Interference Resistance
Safety distances (example) to portable and mobile
radio communication equipment:
*
PEIRP is the “equivalent isotropic radiated power” of the adjacent RF transmitter
** 3 V/m distance to transmitters with frequencies ranging from 150 kHz to 2.5 GHz, otherwise 1 V/m distance.
Relevant Standards
EN 740, Anesthetic workstations and their modules -
Special Regulations, 1998
IEC 60601-1:1988, +A1:1991, + A2:1995: Medical
electrical equipment - Part 1: General requirements
for safety
IEC 60601-1-2:2001: Medical electrical equipment -
Part 1-2: General requirements for safety - Collateral
standard: Electromagnetic compatibility -
Requirements and tests
ASTM F 1101-90 Standard Specification for
Ventilators Intended for Use During Anesthesia
Max.
PEIRP* (W)
3 V/m
Distance **
(m)
1 V/m
Distance
(m)
Indication
0.001 0.06 0.17
0.003 0.10 0.30
0.010 0.18 0.55
0.030 0.32 0.95 e.g., WLAN 5250/5775 (Europe)
0.100 0.58 1.73 e.g., WLAN 2440 (Europe), Bluetooth
0.200 0.82 2.46 e.g., WLAN 5250 (Outside Europe)
0.250 0.91 2.75 e.g., DECT devices
1.000 1.83 5.48 e.g., E-Netz (German digital cellular telephone
network), UMTS, GSM 1900 cellular telephone,
WLAN 5600 (Outside Europe)
2.000 2.60 7.78 e.g., D-Netz (German digital cellular telephone
network used by two providers) cellular
telephones
3.000 3.16 9.49