Specifications Chapter 13 - Specifications
Apollo Operator’s Manual 213
Part Number: 90 38 237, 6th edition
Ventilator
(electronically controlled, electrically driven piston ventilator, fresh gas decoupled)
* Due to gas measurement sampling and leaks (both at the patient and in the device), the maximum delivered tidal volume may be limited to 1100 mL with high
PEEP settings and long expiratory phases.
** Due to gas measurement sampling and leaks (both at the patient and in the device), the end-expiratory PEEP value may be lower than specified at the end of long
expiratory phases.
Ventilation modes Volume mode, pressure mode, pressure support mode (optional),
synchronized volume and pressure modes
Settings:
Pressure limitation P
MAX in volume mode (PEEP+10 cmH
2
O) to 70 cmH
2
O
Accuracy ±10% of setting or ±3 cmH
2
O, whichever is greater
Inspiration pressure P
INSP in pressure mode (PEEP+5 cmH
2
O) to 70 cmH
2
O
Accuracy ±10% of setting or ±3 cmH
2
O, whichever is greater
Tidal volume V
T (compliance-compensated) in
volume mode
20 to 1400 mL*
with optional pressure support 10 to 1400 mL*
Accuracy (10 to 150 mL) ±10% of setting or ±10 mL, whichever is greater
Accuracy (over 150 mL) ±5% of setting or ±15 mL, whichever is greater
Frequency 3 to 80 bpm
Accuracy ±10% of setting or ±1 bpm, whichever is lower
T
INSP 0.2 to 6.7 seconds
Insp./exp. time ratio I:E max. 5:1
Inspiration pause T
IP : TINSP 0 to 60%
Inspiration flow Derived from V
T and TINSP
in volume mode 0.1 to 100 L/min ±10%
in pressure mode max. 150 L/min +10%
PEEP
in volume mode 0 to 20 cmH
2
O (max. PMAX - 10 cmH
2
O)
in pressure mode 0 to 20 cmH
2
O (max. PINSP - 5 cmH
2
O)
Accuracy ±10% of setting or ±2 cmH
2
O, whichever is greater**
∆P
PS 3 to 50 cmH
2
O (max. PINSP)
Trigger 0.3 to 15 L/min
T
SLOPE, in pressure mode and pressure support
mode
0 to 2 seconds