Specifications Chapter 13 - Specifications
Apollo Operator’s Manual 215
Part Number: 90 38 237, 6th edition
Frequency Measurement
Flow measurement
(hot wire anemometry)
Flow
Range –180 to 180 L/min
Resolution of the measurement 0.1 L/min
Accuracy at 60 L/min ±8% of measured value
Tidal volume V
T
Range 0 to 9999 mL
Resolution of the indication 1 mL
Accuracy ±8% of the measured value or ±5 mL, whichever is greater
Volume V
Tinsp
Range 0 to 9999 mL
Resolution of the indication 1 mL
Accuracy ±8% of the measured value or ±5 mL, whichever is greater
Minute volume MV
Range 0 to 99.9 L/min
Resolution of the indication 0.1 L/min
Accuracy ±8% of the measured value or ±0.1 L/min, whichever is greater
Compliance C
PAT
Range 0 to 250 mL/cmH
2
O
Resolution of the indication 0.1 mL/cmH
2
O
Accuracy ±15% of the measured value or ±0.5 mL/cmH
2
O, whichever is greater
MV
LEAK
Range 0 to 9.99 L/min
Resolution of the indication 0.01 L/min
Accuracy ±15% of (MV
EXP + MVLEAK) or ±0.01 L/min, whichever is greater
Frequency Freq./spont. Freq.
Range 1 to 100 bpm
Resolution of the indication 1 bpm
Accuracy ±10% or ±1 bpm, whichever is lower (6 to 100 bpm);
±0.3 bpm (<6 bpm)