EasyManuals Logo

Dräger Apollo Operating Instructions

Dräger Apollo
232 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #223 background imageLoading...
Page #223 background image
Specifications Chapter 13 - Specifications
Apollo Operator’s Manual 215
Part Number: 90 38 237, 6th edition
Frequency Measurement
Flow measurement
(hot wire anemometry)
Flow
Range –180 to 180 L/min
Resolution of the measurement 0.1 L/min
Accuracy at 60 L/min ±8% of measured value
Tidal volume V
T
Range 0 to 9999 mL
Resolution of the indication 1 mL
Accuracy ±8% of the measured value or ±5 mL, whichever is greater
Volume V
Tinsp
Range 0 to 9999 mL
Resolution of the indication 1 mL
Accuracy ±8% of the measured value or ±5 mL, whichever is greater
Minute volume MV
Range 0 to 99.9 L/min
Resolution of the indication 0.1 L/min
Accuracy ±8% of the measured value or ±0.1 L/min, whichever is greater
Compliance C
PAT
Range 0 to 250 mL/cmH
2
O
Resolution of the indication 0.1 mL/cmH
2
O
Accuracy ±15% of the measured value or ±0.5 mL/cmH
2
O, whichever is greater
MV
LEAK
Range 0 to 9.99 L/min
Resolution of the indication 0.01 L/min
Accuracy ±15% of (MV
EXP + MVLEAK) or ±0.01 L/min, whichever is greater
Frequency Freq./spont. Freq.
Range 1 to 100 bpm
Resolution of the indication 1 bpm
Accuracy ±10% or ±1 bpm, whichever is lower (6 to 100 bpm);
±0.3 bpm (<6 bpm)

Table of Contents

Other manuals for Dräger Apollo

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Apollo and is the answer not in the manual?

Dräger Apollo Specifications

General IconGeneral
BrandDräger
ModelApollo
CategoryMedical Equipment
LanguageEnglish

Related product manuals