Ventilator and Fresh Gas Delivery Failure Chapter 12 - Troubleshooting
Apollo Operator’s Manual 195
Part Number: 90 38 237, 5th edition
Ventilator and Fresh Gas Delivery
Failure
Figure 163. Ventilator and Gas Mixer Failure Messages
If both the ventilator and gas mixer fail, the following
message is displayed in the status field on the
screen: “GAS + VENT FAIL”.
The ventilation soft keys are removed from the
screen and a prompt appears advising the user how
to proceed:
Ventilator failure!
Only manual ventilation possible.
The machine automatically switches to Monitoring
mode.
Ventilate the patient manually
Note: Refer to the total flow meter for approximate
flow (see the Specifications section of the
manual for accuracy).