Configuring the Default Settings in Standby Chapter 10 - Configuration
Apollo Operator’s Manual 155
Part Number: 90 38 237, 6th edition
Adjustment Ranges for Default Alarm Limits
Two dashes (--) indicate that the alarm is disabled.
Alarm trigger times are as follows:
All apnea, apnea pressure, apnea volume, and
apnea CO
2
alarms are active after 35 seconds in the
mechanical ventilation modes with a frequency of
less than 6 bpm and in pressure support mode with a
minimum frequency (Freq
MIN) of less than 6 bpm or
when set to OFF.
Certain alarms can be disabled automatically in Man/
Spont, Monitoring mode, and Aux. CGO mode
(optional) (see page 153).
Alarm
Adjustment
range
Factory Default
Value
SpO
2
[%]
>
<
51 to 99, --
50 to 98
--
92
Pulse
[bpm]
>
<
21 to 250
20 to 249
120
50
etCO2
[mmHg]
>
<
1 to 75, --
0 to 74, --
50
--
MV
[L/min]
>
<
0.1 to 20.0, --
0 to 19.9, --
12
3.0
FiO2
[Vol.%]
>
<
19 to 99, --
18 to 98
--
20
in Hal.
[Vol.%]
>
<
0.1 to 8.4
0 to 8.3, --
1.5
--
in Iso.
[Vol.%]
>
<
0.1 to 8.4
0 to 8.3, --
2.3
--
in Enf.
[Vol.%]
>
<
0.1 to 9.9
0 to 9.8, --
3.4
--
in Des.
[Vol.%]
>
<
0.1 to 21.9
0 to 21.8, --
12.0
--
in Sev.
[Vol.%]
>
<
0.1 to 9.9
0 to 9.8, --
4.2
--
PAW
[cmH
2
O]
>
<
5 to 99
0 to 35
40
8
Apnea pressure after 20 seconds
Apnea flow after 20 seconds
Apnea CO
2
after 20 seconds
(after 65 seconds in Man/Spont,
monitoring mode, and Aux. CGO
mode)