Chapter 11 - Cleaning and Maintenance Care List for Apollo Components
178 Apollo Operator’s Manual
Part Number: 90 38 237, 6th edition
Care List for Apollo Components
The following table lists Apollo components with
recommended processing methods. Processing refers
to cleaning, disinfection, and/or sterilization, as
appropriate for a given component.
The table is intended as a guide. Follow the institution’s
policies regarding specific methods and agents for
cleaning and sterilization.
The parts should be cleaned and disinfected by
machine. If not, they must be disinfected by immersion
and then cleaned.
CAUTION !
Apollo and its components must not be treated with
formaldehyde vapors or ethylene oxide.
Component Processing Method
Disinfection and cleaning Sterilization
in steam
Cleaning/
disinfecting
machine
1
93
o
C (199
o
F)
10 minutes
Wiping
2
Disinfection
by immersion
2
)
134
o
C (273
o
F)
8 minutes
Apollo workstation No Outside
3
No No
Power cable, connecting hoses No Yes No No
Breathing bag arm Yes Yes Yes
8
Yes
Ventilator diaphragm
4
Yes No Yes Yes
Cover of breathing system
with APL valve
Yes
5
No Yes Yes
6
Middle and bottom part of breathing
system
Yes
5
No Yes Yes
6
Expiratory port/inspiratory port Yes No Yes Yes
Absorber and insert Yes No Yes Yes
Spirolog flow sensors No No Yes
7
No
SpiroLife flow sensors No No Yes
7
Yes
AGS housing Yes Yes Yes No
AGS flow tube (without filter) No Yes No No
Container for AGS buffer volume Yes Yes Yes No
Transfer hose (connected to scavenger
port on back of machine)
Yes Yes Yes No
Suction system regulator and bottle
assembly
(Follow manufacturer’s instructions)
1. Only use neutral cleaning agents. Do not use disinfectants due to risk of corrosion.
2. Use disinfectants based on aldehydes and quaternary ammonium compounds, see page 175.
Note: Testing was performed with Incidin Extra N and Incidur (wiping), and Gigasept FF and Korsolex Extra (disinfection by immersion).
3. Do not use any agents containing alcohol.
4. Remove any water which may have accumulated in the ventilator diaphragm.
Larger quantities of condensation may impair operation of the workstation and/or lead to failure of the equipment.
5. Only with fully demineralized water.
6. The valve plate must be sterilized after washing in order to dry it. Correct operation of the workstation may be impaired and lead to failure of the
workstation if the control areas in the valve plate are not dried completely.
7. Disinfect the flow sensor in 70% ethanol or isopropanol solution for approx. 1 hour and leave to dry in air for at least 30 minutes.
8. Use disinfectants based on glutaraldehyde and ortho phthaldehyde compounds.
Note: testing was performed with Cidex and Cidex OPA (disinfection by immersion).

Table of Contents

Other manuals for Dräger Apollo

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Apollo and is the answer not in the manual?

Dräger Apollo Specifications

General IconGeneral
BrandDräger
ModelApollo
CategoryMedical Equipment
LanguageEnglish

Related product manuals