EasyManuals Logo

Dräger Apollo Operating Instructions

Dräger Apollo
232 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #185 background imageLoading...
Page #185 background image
Cleaning and Disinfection Guidelines Chapter 11 - Cleaning and Maintenance
Apollo Operator’s Manual 177
Part Number: 90 38 237, 6th edition
Do not clean the SpiroLife flow sensor in a
cleaning and disinfection machine or with
compressed air, water jets, brushes, etc. These
methods would damage the thin sensor wires
inside.
The flow sensor can be reused as long as it can
be calibrated successfully.
The flow sensor is not suitable for plasma or
radiation sterilization.
CLIC Adapter
In the event of major condensation in the adapter (e.g.
during minimum flow operation), disinfect the adapter
in moist heat together with the disassembled breathing
system after not more than 24 hours of operation. This
will ensure that the adapter is cleaned, disinfected, and
dried.
WARNING !
If condensation flows back from the breathing
tubes into the adapter (tubes routed upwards),
the breathing tubes must be regularly drained at
short intervals.

Table of Contents

Other manuals for Dräger Apollo

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Apollo and is the answer not in the manual?

Dräger Apollo Specifications

General IconGeneral
BrandDräger
ModelApollo
CategoryMedical Equipment
LanguageEnglish

Related product manuals