Chapter 11 - Cleaning and Maintenance Apollo Cleaning and Disinfection
168 Apollo Operator’s Manual
Part Number: 90 38 237, 6th edition
Removing the patient system hoses
Figure 134. Patient System Hoses
• Disconnect the breathing hoses from the
breathing system (1 in Figure 134).
• Disconnect the various parts of the hose
system, including breathing hoses, Y-piece,
connector, and optional Y-piece filter.
• Disconnect the breathing bag and arm (2 in
Figure 134) by loosening the two thumb
screws.
• Prepare the parts for cleaning and disinfection.
Figure 135. Proper Breathing Hose Removal
1
2
1
2
CAUTION !
Do not damage the breathing hoses when removing
them. Always hold the hoses by the connection
sleeve and not by the spiral ribbing when removing
and connecting the breathing hoses, otherwise the
spiral ribbing may be torn off the sleeve. Breathing
hoses with damaged spiral ribbing can be easily
kinked which can interrupt the gas flow.
Always check the breathing hoses for damage prior
to use. Damaged breathing hoses must not be
used.