Apollo Cleaning and Disinfection Chapter 11 - Cleaning and Maintenance
Apollo Operator’s Manual 167
Part Number: 90 38 237, 6th edition
Apollo Cleaning and Disinfection
This section of the manual provides complete
instructions for the disassembly and cleaning of the
Apollo anesthesia machine.
Note: Set the Apollo to Standby before
disassembly.
Figure 132. Removing the Sample Line
Disassembling Components
Removing the sample line
Disconnect the sample line from the Y-piece
and the fitting on the water trap (see Figure
132).
Figure 133. Removing the Water TrapRemoving the water trap
Pull the water trap straight out of its holder.
CAUTION !
Sample gas lines are single-use articles and must
not be disinfected.
Disinfectants can damage the sample gas line and
the diaphragm of the water trap.
WARNING !
The water trap must be replaced after a service
life of four weeks.
The diaphragm could become brittle after an
extended period of use, allowing water and
bacteria to get into the measurement system.
The contamination affects the gas
measurement, or the gas measurement may fail
as a result.
CAUTION !
Do not allow any alcohol or cleaning agents/
disinfectants to come into contact with the inside of
the water trap.
Do not wash, flush or sterilize the water trap.
These substances can damage the diaphragm and
the measurement system.

Table of Contents

Other manuals for Dräger Apollo

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Apollo and is the answer not in the manual?

Dräger Apollo Specifications

General IconGeneral
BrandDräger
ModelApollo
CategoryMedical Equipment
LanguageEnglish

Related product manuals