Chapter 9 - Alarms Displaying and Setting Alarm Limits
142 Apollo Operator’s Manual
Part Number: 90 38 237, 6th edition
Adjustment range of the alarm limits during operation
To exit the alarm limits menu:
confirm the exit symbol “z“ with the
confirm knob, or
press the > < key.
Alarm
Adjustment
range
SpO
2
[%]
>
<
51 to 99, --
50 to 98
Pulse
[bpm]
>
<
21 to 250
20 to 249
etCO2
[mmHg]
>
<
1 to 75, --
0 to 74, --
FiCO2
[mmHg]
> 1 to 10, --
MV
[L/min]
>
<
0.1 to 20.0, --
0 to 19.9, --
FiO2
[Vol.%]
>
<
19 to 99, --
18 to 98
in Hal.
[Vol.%]
>
<
0.1 to 8.4
0 to 8.3, --
in Iso.
[Vol.%]
>
<
0.1 to 8.4
0 to 8.3, --
in Enf.
[Vol.%]
>
<
0.1 to 9.9
0 to 9.8, --
in Des.
[Vol.%]
>
<
0.1 to 21.9
0 to 21.8, --
in Sev.
[Vol.%]
>
<
0.1 to 9.9
0 to 9.8, --
PAW
[cmH
2
O]
>
<
5 to 99
0 to 35