If the mean of the last (up to four) measured breaths
cannot be calculated, the measured value of the last
breath is used instead.
Volume mode,
pressure mode,
pressure support mode
PAW
high alarm limit
[cmH
2
O]
PEAK + 5 cmH
2
O
or
PLAT + 10 cmH
2
O,
whichever is greater
P
AW
low alarm limit
[cmH
2
O]
(PLAT – PEEP) x 0.6
+ PEEP – 1,
but at least
3 cmH
2
O in
vol. mode/press. mode/
press. supp. mode