Chapter 10 - Configuration Configuring the Default Settings in Standby
158 Apollo Operator’s Manual
Part Number: 90 38 237, 6th edition
- PAW low alarm limit adjusts with PEEP: (Yes or No)
If this setting is set to “Yes”, the low alarm limit for
airway pressure (PAW) will be automatically changed
when the PEEP value is changed.
If this setting is set to “No”, the low alarm limit for
airway pressure (PAW) will be unaffected by changes
in the PEEP value.
In the Pressure Mode, the PAW low alarm limit will
not exceed the PINSP - 2. This is also true for
changes to PINSP.
In the Pressure Support Mode, PEEP + ∆P
PS - 2 will
not be exceeded. This is also true for changes to
∆P
PS.