Chapter 12 - Troubleshooting Alarm – Cause – Remedy
206 Apollo Operator’s Manual
Part Number: 90 38 237, 5th edition
Condition Cause Remedy
“INOP” displayed instead of
measured values
Values cannot be measured, sensor
faulty.
Replace sensor if necessary.
Ensure adequate substitute monitoring.
Call DrägerService.
“CAL” displayed instead of
measured values
Sensors are being calibrated. Wait until calibration is complete.
““ beside measured
values
All alarms for the measured values
concerned
1
have been disabled.
Enable alarms in configuration menu (see
page 141).
““ displayed beside O
2
label
O
2
alarm is inactive because no valid
CO
2
breath detected.
Check sample line. Wait until CO
2
waveform has been detected.
Symbol , , or
beside measured values
One or both alarm limits for the measured
value concerned has/have been
disabled.
Set alarm limits (see page 141).
Grayed out measured values The set value differs from the delivered
value.
The specified accuracy cannot be
maintained.
“- -” (dashes) displayed
instead of measured values
Alarm limit is disabled. Set alarm limits (see page 141).
1. Applies for all CO2 and SpO2 alarms.
g
_
>
<

Table of Contents

Other manuals for Dräger Apollo

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Apollo and is the answer not in the manual?

Dräger Apollo Specifications

General IconGeneral
BrandDräger
ModelApollo
CategoryMedical Equipment
LanguageEnglish

Related product manuals