Emergency Start Chapter 5 - Preuse Checkout
Apollo Operator’s Manual 69
Part Number: 90 38 237, 6th edition
Note: The >Cancel Test< key is also available in
the self test screen.
WARNING !
The leak and compliance tests are not
performed if the self test is cancelled, and no
leak and compliance information is available.
The accuracy levels specified in the
Specifications chapter of this manual cannot be
guaranteed.
CAUTION !
Cancelling the self test may lead to limited
functionality. Greater clinical attention is required
during operation if the self test has been
interrupted. The clinician should perform a complete
power-up self test as soon as is practical.