Chapter 6 - Operation Summary Changing Patients
78 Apollo Operator’s Manual
Part Number: 90 38 237, 6th edition
3. Mount the new disposable absorber as follows:
• Before fitting, shake the
disposable absorber, e.g. by
turning it upside down several
times in order to loosen up the
soda lime.
• Remove seal from new
disposable absorber.
• Slide the new disposable
absorber onto the mounting and
4. Push the absorber into the machine until it
engages.
5. After ending the case, perform a leak test to
determine leak and compliance data.
6. Reset the absorbent change log to actual date
(see page 64).