Chapter 7 - Ventilation Manual/Spontaneous Ventilation
88 Apollo Operator’s Manual
Part Number: 90 38 237, 6th edition
If a Man/Spont parameter has to be changed
during ventilation:
Press the soft key for the
parameter to be changed, turn
the confirm knob to the desired
value, and press knob to
confirm.
Note: Man/Spont mode can also be started without
presetting values:
Press the >Man Spont< key
located at the bottom of the
display panel, and press the
confirm knob. Default parameter
values are used.
or
If in Standby or Monitoring
mode, set fresh-gas flow. This
activates manual/spontaneous
ventilation automatically. Default
parameter values are used.
Certain alarms are disabled automatically in manual/
spontaneous mode to avoid artifacts. See the
“Alarms” chapter in this manual for a list of alarms
active in manual/spontaneous mode.
Note: There is 15-second timeout period for
making ventilation mode changes, with a
5-second audible tone sequence after the
first 10 seconds. If the new setting is not
confirmed within the timeout period, the
current ventilation setting remains in effect.

Table of Contents

Other manuals for Dräger Apollo

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Apollo and is the answer not in the manual?

Dräger Apollo Specifications

General IconGeneral
BrandDräger
ModelApollo
CategoryMedical Equipment
LanguageEnglish

Related product manuals