Chapter 7 - Ventilation Manual/Spontaneous Ventilation
88 Apollo Operator’s Manual
Part Number: 90 38 237, 6th edition
If a Man/Spont parameter has to be changed
during ventilation:
Press the soft key for the
parameter to be changed, turn
the confirm knob to the desired
value, and press knob to
confirm.
Note: Man/Spont mode can also be started without
presetting values:
Press the >Man Spont< key
located at the bottom of the
display panel, and press the
confirm knob. Default parameter
values are used.
or
If in Standby or Monitoring
mode, set fresh-gas flow. This
activates manual/spontaneous
ventilation automatically. Default
parameter values are used.
Certain alarms are disabled automatically in manual/
spontaneous mode to avoid artifacts. See the
“Alarms” chapter in this manual for a list of alarms
active in manual/spontaneous mode.
Note: There is 15-second timeout period for
making ventilation mode changes, with a
5-second audible tone sequence after the
first 10 seconds. If the new setting is not
confirmed within the timeout period, the
current ventilation setting remains in effect.