Instructions for use Primus Infinity Empowered SW 4.5n 113
Operation
Pressure sup-
port
PPS
5)
[hPa (cmH2O)]
OFF,
3 to 50 5
(Pressure
Support)
max. P
MAX
PEEP
OFF
(Volume
Mode/
Pressure
Mode)
Rise time
TSLOPE
[sec.]
0.0 to 2.0 0.0
age [years] <1 to 120 40
weight
[kg]
[lbs]
1 to 120
1 to 240
– –
1) The default values can be set specifically for the hospital
concerned, see page 166.
2) Depending on the configuration, the inspiratory time
TINSP can be automatically changed together with adjust-
ment of the frequency so that the resultant ratio of inspi-
ration to expiration I:E remains constant. Only applies if
trigger = OFF, see "Ventilator and gas delivery"
on page 175.
3) The resultant ratio of inspiration to expiration I:E is also
displayed in parallel.
4) Depending on the configuration, the pressure limit value
PINSP can also be changed automatically when the PEEP
value is changed. See "Starting the preset ventilation
mode" on page 116 and "Ventilator and gas delivery"
on page 175.
5) Optional
Ventilation pa-
rameters
Setting
ranges
Factory set-
ting
1)