Operation
112
Instructions for use Primus Infinity Empowered SW 4.5n
Synchronization is maintained with the set value
when pressure support is deactivated and set to
OFF.
In case of a continuous and strong patient activity,
it is possible that the mandatory ventilation coin-
cides with the pressure supported patient breaths,
resulting in an increased tidal volume V
T.
Pressure support is automatically deactivated
when the trigger is deactivated and set to OFF.
The current trigger status is shown above the keys
for the ventilation parameters (H).
Setting ranges and factory settings
Ventilation pa-
rameters
Setting
ranges
Factory set-
ting
1)
Pressure limita-
tion
PINSP
[hPa (cmH2O)]
5 to 70
min. PEEP +5
15
Frequency
freq.
2) 3)
[1/min]
3 to 100 12
Inspiratory time
TINSP
3)
[sec.]
0.2 to 6.7 1.7
Insp. pause time:
Insp. time
TIP:TINSP [%]
0 to 60 10
PEEP
4)
[hPa (cmH2O)]
0 to 20
max. P
INSP –5
0
Trigger sensitivity
Trigger
[L/min]
OFF,
0.3 to 15
3.0
(Pressure
Support)
OFF
(Volume
Mode/
Pressure
Mode)