Instructions for use Primus Infinity Empowered SW 4.5n 195
Remedying faults
Caution/
Note
(24/7)
AIR CYLIND. EMPTY * AIR reserve gas cylinder
empty and central AIR sup-
ply not available or not con-
nected.
Use a new AIR reserve gas
cylinder.
Use the central gas supply.
Caution
(24)
Warning
(31)
APNEA Priority in accordance with
maximum priority of the in-
dividual alarms.
Breathing/ventilation has
stopped (detected by pres-
sure, volume, and CO
2
monitoring).
Patient must immediately
be ventilated manually!
Check patient's spontane-
ous breathing ability.
Check ventilation setting.
Check fresh-gas setting.
Make sure everything is
connected.
Check hose system and
tube!
Priority Message Cause Remedy