Remedying faults
196
Instructions for use Primus Infinity Empowered SW 4.5n
Note
(10)
APNEA CO2 Apnea alarms are graded
in time:
In automatic ventilation
modes:
Caution = 0 to 30 sec.
Warning = >30 sec.
In ventilation modes
Man.Spont., Pressure
Support, Ext. Outlet:
Note = 0 to 30 sec.
Caution = 31 to 60 sec.
Warning = >60 sec.
Sample line not connected. Check sample line.
Caution
(24)
Warning
(31)
No spontaneous breathing. Patient must immediately
be ventilated manually!
Check patient's spontane-
ous breathing ability.
Make sure everything is
connected.
Check hose system and
tube.
Breathing/ventilation has
stopped.
Patient must immediately
be ventilated manually!
Check ventilation setting.
Priority Message Cause Remedy