Instructions for use Primus Infinity Empowered SW 4.5n 197
Remedying faults
Caution/
(see
page 174)
Note
(
see
page 181)
(10)
Caution
(24)
= 0 to
30 sec.
APNEA FLOW Breathing/ventilation has
stopped.
Patient must immediately
be ventilated manually!
Warning
(31)
>30 sec.
Check patient's spontane-
ous breathing ability.
Check ventilation setting.
Insufficient fresh gas sup-
ply.
Check fresh-gas setting.
Tube kinked.
Leak in hose system.
Check hose system and
tube.
Caution
(24)
= 0 to
30 sec.
APNEA PRESSURE Breathing/ventilation has
stopped
Patient must immediately
be ventilated manually!
Warning
(see
page 174)
(31)
≥0
or
>30 sec.
Insufficient fresh gas sup-
ply.
Check fresh-gas setting.
Leak or blockage in tube or
hose system.
Check hose system, tube,
and microbial filter.
Patient not connected. Connect patient correctly.
Priority Message Cause Remedy