Instructions for use Primus Infinity Empowered SW 4.5n 201
Remedying faults
Note
(1)
GAS SENSOR FAIL
(in Standby only)
Complete gas measure-
ment system failure.
Use external gas measur-
ing system.
Call DrägerService.
Note
(7)
HOSE EXPIRED
(Only available if
ID ventilation circuit used
has been activated in the default
configuration.)
Breathing hose shelf life
has been expired.
Exchange breathing hoses.
Caution/
–
HOSE MISMATCH * Breathing hoses are not
connected to the correct
breathing system port.
Check breathing hose con-
nection for mismatch.
(11/–) Wrong hose type con-
nected to a breathing sys-
tem port.
Check type of breathing
hoses.
Unknown hose or acces-
sory connected.
Remove unknown acces-
sories.
Check whether breathing
system is correctly in-
stalled.
Note
(7)
ID FUNCTIONS FAIL
(Not available if the Infinity ID func-
tion has been completely deacti-
vated by DrägerService.)
Infinity ID functions not
available, because of a de-
vice failure.
Call DrägerService.
Note
(7)
ID HOSE MISSING
(Only available if ID ventilation
circuit used is activated in the de-
fault configuration.)
No or not all Infinity ID
breathing hoses are de-
tected.
Connect all Infinity ID
breathing hoses correctly.
Caution/
–
(11/–)
INCOMP. HOSE * Breathing hoses and other
accessories, which are
connected, are not in-
tended to be used with
Primus Infinity
Empowered.
Exchange breathing hoses.
Use only compatible acces-
sories with Primus Infinity
Empowered.
Priority Message Cause Remedy