Remedying faults
202
Instructions for use Primus Infinity Empowered SW 4.5n
Caution/
Note
(14/10)
INCORRECT FG FLOW * Set fresh-gas flow cannot
be delivered.
Check vaporizer setting.
Make sure the emergency
O
2 delivery is closed.
Set fresh-gas flow to be-
tween 3 and 10 L/min.
Caution
(11)
INSP. CO2 HIGH Soda lime in circle system
exhausted.
Increase the fresh-gas flow.
Replace soda lime.
Leak or fault in breathing
system.
Replace breathing system!
High respiratory rates.
If the respiratory rates are
high, the measured value
can no longer follow the
gas concentration properly
due to the design of the
system.
Adjust alarm limits if neces-
sary.
Dead space ventilation. Check ventilation settings.
Note
(8)
INSP. FLOW SENSOR FAIL Insp. flow sensor is defec-
tive.
Replace flow sensor,
see page 220.
Priority Message Cause Remedy