Instructions for use Primus Infinity Empowered SW 4.5n 203
Remedying faults
Caution
Warning
Note
INSP. HAL. HIGH
INSP. ISO. HIGH
INSP. ENF. HIGH
INSP. DES. HIGH
INSP. SEV. HIGH
Caution (24) =
insp. MAC value >3 MAC
for >180 seconds.
Warning (31) =
insp. MAC value >5 MAC
Warning (31) =
insp. MAC value >3 MAC
and exp. MAC value
>2.5 MAC for
>30 seconds.
Insp. anesthetic gas con-
centration exceeds 5 MAC.
Check the vaporizer and
fresh-gas settings.
Insp. anesthetic gas con-
centration exceeds 3 MAC
for more than 180 seconds.
Insp. anesthetic gas con-
centration exceeds 3 MAC
and exp. 2.5 MAC for more
than 30 seconds.
Note (10) =
Insp. gas concentration >
upper alarm limit for 0 to
30 seconds.
(preliminary information for
the user)
Caution (24) =
Insp. gas concentration >
upper alarm limit for 31 to
180 seconds.
Warning (31) =
Insp. gas concentration >
upper alarm limit for
>180 seconds.
Inspiratory anesthetic gas
concentration exceeds the
upper alarm limit for at least
2 breaths.
Check the vaporizer and
fresh-gas settings.
Priority Message Cause Remedy