Instructions for use Primus Infinity Empowered SW 4.5n 209
Remedying faults
Note
(1)
POWER SPLY. ERROR Internal fault in the power
supply.
Call DrägerService.
Operation of the worksta-
tion can continue for the
time being.
Note
(8)
PRESSURE ERROR
(in Standby only)
Pressure sensor is defec-
tive.
Perform self test.
Call DrägerService.
Warning
(27)
PRESSURE HIGH Upper alarm limit for the
airway pressure has been
exceeded.
Tube kinked. Check hose system and
tube.
Stenosis.
Ventilation settings not cor-
rect.
Correct ventilation settings.
Caution
(13)
PRESSURE LIMITATION Ventilator is operating with
pressure limitation.
Check ventilation setting.
Tube kinked/stenosis. Check tube, hoses, and fil-
ter.
Microbial filter contami-
nated on inspiration side.
Check microbial filter.
Warning
(30)
PRESSURE NEGATIVE Insufficient fresh gas sup-
ply.
Set adequate fresh-gas
flow on anesthetic ma-
chine. Flush system if nec-
essary.
Endotracheal aspiration
during ventilation.
Check endotracheal aspi-
ration system.
Negative pressure due to
fault in ventilator.
Make sure upper dia-
phragm is correctly in-
stalled.
Call DrägerService.
Anesthetic gas scavenging
system is defective.
Check anesthetic gas scav-
enging system.
Call DrägerService.
Priority Message Cause Remedy