Remedying faults
210
Instructions for use Primus Infinity Empowered SW 4.5n
Note
(10)
PRESSURE RELIEF Internal pressure relief
valve opened due to high
system pressure.
(Standby, Monitoring,
Man.Spont.)
Check APL valve settings.
Check fresh-gas settings.
Caution
(21)
PULSE RATE HIGH Upper alarm limit for pulse
has been exceeded.
Check patient's condition!
Correct alarm limit if neces-
sary.
Warning
(31)
PULSE RATE LOW Pulse below lower alarm
limit.
Check patient's condition!
Check ventilation.
Caution
(18)
SAFETY O
2 OPEN O2 emergency delivery is
open.
Close O2 emergency sup-
ply.
Caution/
–
(14/–)
SETTING CANCELLED * The last settings have not
been accepted on account
of temporary errors.
Repeat settings.
Alarm can be canceled by
pressing .
Note
(7)
SODA LIME DEPLETED?
(in Standby only)
Soda lime of Infinity ID
CLIC Absorber in circle
system exhausted.
(Only available if this func-
tionality is activated by
DrägerService.)
Check soda lime, if neces-
sary replace soda lime.
Time limit for usage
reached.
Shelf life of Infinity ID CLIC
Absorber expired.
(Only available if this func-
tionality is activated by
DrägerService.)
Caution
–
(11/–)
SODA LIME DISCONN.* Infinity ID CLIC Absorber
not correctly connected.
Check and connect
Infinity ID CLIC Absorber.
Priority Message Cause Remedy