Remedying faults
212
Instructions for use Primus Infinity Empowered SW 4.5n
Caution
(12)
VOLUME NOT ATTAINED Set volume is not delivered. Remove leak.
Correct pressure limitation
or inspiratory time if neces-
sary. Check fresh-gas flow
setting.
Check fresh-gas flow set-
ting.
Note
(7)
WATER TRAP EXPIRED
(This alarm is only available if this
functionality is activated by
DrägerService.)
Service life of Infinity ID
water trap for gas analyzer
expired.
Exchange water trap.
Note
(7)
WATER TRAP SAMPL.
LINE?
Sample line blocked or not
connected. Water trap or
gas measurement system
blocked or not connected.
Check sample line, water
trap, gas measurement
system, and filter in
Y-piece, if applicable; re-
place if necessary.
INOP instead of measured
values
Values cannot be meas-
ured, sensor defective.
Replace sensor if neces-
sary.
Ensure adequate substi-
tute monitoring!
Call DrägerService.
CAL instead of measured
values
Sensors are being cali-
brated.
Wait until calibration is
complete.
"– –" instead of measured
values
Measurement currently not
possible.
Ensure adequate substi-
tute monitoring!
Call DrägerService.
symbol next to meas-
ured values
All CO
2 and SpO2 alarms
for the measured values
concerned have been disa-
bled.
Enable alarms in configura-
tion menu, see page 171.
symbol next to meas-
ured values
All alarms for the measured
values concerned have
been temporarily disabled.
The alarm system is wait-
ing for automatic measure-
ment wake-up
(AutoWakeUp).
Connect sample line.
Connect SpO
2 Sensor.
Connect patient.
For more details
see page 130.
Priority Message Cause Remedy